U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CRYOLIFE, INC. BIOGLUE SYRINGE 5-PACK, 10 ML, JAPAN; GLUE,SURGICAL,ARTERIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CRYOLIFE, INC. BIOGLUE SYRINGE 5-PACK, 10 ML, JAPAN; GLUE,SURGICAL,ARTERIES Back to Search Results
Model Number BG3510-5-J
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Aneurysm (1708)
Event Date 05/21/2019
Event Type  Injury  
Manufacturer Narrative
This investigation is currently ongoing.Any additional information will be provided in the follow-up report.
 
Event Description
According to the initial report, "the false aneurysm was observed on the central posterior wall, so the repairment procedure was performed on (b)(6) 2019.On (b)(6) 2019, the patient underwent surgery for ascending aorta repairment and aortic valve repairment.For both repairments, bioglue was used on anastomotic part of artificial vessel for supporting hemostasis.There was false aneurysm but the relationship between bioglue and developing of false aneurysm is unclear.".
 
Manufacturer Narrative
The manufacturing records for lots 18mjx008, 18ejx001 and 18ejx002 were reviewed and it was confirmed that all records were controlled, available for review, and met all specifications per the device master record.All lots passed functional testing and met release specifications.The following information is unknown: the condition of the native tissue before surgery, how much bioglue was used in each procedure, or if the syringe was primed and de-aired.Per kitamura et al, pseudoaneurysm formation ¿is not a rare late complication late after repair of acute aortic dissection (mohammadi, s et al).The underlying mechanism of pseudoaneurysm formation is considered to be associated with tissue cutting due to the fragility of the dissected aortic wall at the anastomosis and from the chemical reaction to the aldehyde contained in the glue material (bingley, ja et al).¿ perhaps the native tissue was too damaged to be repaired and an aortic replacement with a synthetic graft should have been considered.Furthermore, while surgical glue is helpful in surgery for acute type a aortic dissection, it may also cause late pseudoaneurysm formation or valve deterioration when not used properly (kitamura et al).Dr.Fehrenbacher et al.Performed a retrospective review of 92 consecutive patients who underwent complex operation in which bioglue was used.Postoperative pseudoaneurysm formation occurred in 3.3% of the patients (fehrenbacher 2006).Weiner et al.Presented at 15th world congress of heart disease in vancouver, canada in july 2010 they identified 97consecutive patients in whom bioglue was used to reinforce thoracic aortic suture lines.During follow-up 2 patients were identified as having a pseudoaneurysm, the control group, without bioglue use, had similar incidences of pseudoaneurysm formation (weiner 2010).Ma et al.Reviewed 233 patients with a mean follow-up time of 2.4 years post-operation; a pseudoaneurysm was detected in only 1 patient (0.6%).The authors concluded, ¿the use of bioglue in thoracic aortic surgery was not associated with excess incidence of anastomotic pseudoaneurysm formation following surgical repair of thoracic aortic disease.¿ (ma 2017) there is insufficient information to determine if there is an association between the use of bioglue and the pseudoaneurysm formed.Pseudoaneurysm formation is a known complication in standard surgical repair of aortic dissections.The condition of the native aortic tissue at the time of initial surgery is unknown in this case.Pre-existing conditions such as aortic medial necrosis or other intrinsic aortic disease may have contributed to further complications.This report is being submitted as required by federal regulations and does not constitute an admission that the device caused or contributed to the reported event.Furthermore, this report reflects the event as alleged by the complainant and does not imply that the information reported to cryolife is accurate or has been confirmed by cryolife.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOGLUE SYRINGE 5-PACK, 10 ML, JAPAN
Type of Device
GLUE,SURGICAL,ARTERIES
Manufacturer (Section D)
CRYOLIFE, INC.
1655 roberts blvd. nw
kennesaw GA 30144
MDR Report Key9284149
MDR Text Key185002794
Report Number1063481-2019-00069
Device Sequence Number1
Product Code MUQ
Combination Product (y/n)N
PMA/PMN Number
P010003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 01/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberBG3510-5-J
Device Lot Number18MJX008
Was Device Available for Evaluation? No
Distributor Facility Aware Date10/07/2019
Initial Date Manufacturer Received 10/07/2019
Initial Date FDA Received11/06/2019
Supplement Dates Manufacturer Received10/07/2019
Supplement Dates FDA Received01/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-