Model Number NGP680300
Device Problem Smoking (1585)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: date unknown.First name: unknown.(b)(6).The system was evaluated by a field service engineer.The field service found inverter was not working properly.The inverter was replaced.A field service checklist was performed.The unit complied with all factory settings.A review of records related to the device including labeling, complaint trending, and risk documentation will be performed.Upon completion of this review, if there is any further relevant information obtained, a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that smoke came from the monitor.There was no patient involvement reported.No additional information was provided.
 
Manufacturer Narrative
In initial report, the manufacturer year provided was inadvertently reported as 4/26/2008, however the correct date is 4/23/2008.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa: 010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WHITESTAR SIGNATURE SYSTEM
Type of Device
PHACOEFRAGMENTATION UNIT
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key9284223
MDR Text Key165224623
Report Number3006695864-2019-00879
Device Sequence Number1
Product Code HQC
UDI-Device Identifier05050474534476
UDI-Public(01)05050474534476
Combination Product (y/n)Y
PMA/PMN Number
K060366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 01/01/2005,11/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNGP680300
Device Catalogue NumberNGP680300
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 10/15/2019
Initial Date FDA Received11/06/2019
Supplement Dates Manufacturer Received12/12/2019
10/25/2020
Supplement Dates FDA Received01/08/2020
11/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-