Model Number 056
Device Problems Thermal Decomposition of Device (1071); Fire (1245)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
Customer stated, " the cord caught fire by the switch." product was returned for investigation.Customer did not claim injury.Investigator observed a burn cut near the strain relief.The cut was the cause of the spark/ fire the customer saw.Investigator determined that wear and tear of the product over 17 years of constant use caused the issue.Bce states in the medical device files that the product life is four years.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOPHORE
Type of Device
THERMOPHORE
Manufacturer (Section D)
BATTLE CREEK EQUIPMENT
702 s reed rd
fremont IN 46737
Manufacturer Contact
michael green
MDR Report Key9285307
MDR Text Key173283148
Report Number1832415-2019-10828
Device Sequence Number1
Product Code IRT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 11/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number056
Device Lot Number023161
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2019
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/06/2019
Was Device Evaluated by Manufacturer? Yes
Type of Device Usage N
Patient Sequence Number1
-
-