• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG CUSHING RONGEUR STR 2X10MM180MM; BONE PUNCHES, RONGEURS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG CUSHING RONGEUR STR 2X10MM180MM; BONE PUNCHES, RONGEURS Back to Search Results
Model Number FF803R
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with a cushing rongeur.The jaw was broken during surgery on the bone according to the report.There was no patient harm.An additional medical intervention was not necessary.This malfunction prolonged the surgery for 5 minutes.Additional information was not provided nor available.The malfunction is filed under aag reference (b)(4).
 
Manufacturer Narrative
Investigation results: visual investigation: vigilance investigator carried out the pictorial documentation visually and microscopically.During investigation the breakage surfaces of the broken remained part on the instrument has been analysed.The analysis of the fracture pattern illustrated a forced fracture due to overload.No pores, inclusions or foreign bodies could be found on the point of rupture.Furthermore the mouth shows several damages on the cutting edge and its lateral parts.Batch history review: the device quality and manufacturing history records for the available lot number has been requested at the responsible q-coordinator of the production plant.The answer is still pending, the 8d-report will be updated upon receipt of the review-confirmation.No similar incidents have been filed with products from this batch.The review of risk assessment revealed that the overall risk level (severity x probability of occurrence) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a capa is not necessary.
 
Event Description
The malfunction is filed under aag reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSHING RONGEUR STR 2X10MM180MM
Type of Device
BONE PUNCHES, RONGEURS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key9286210
MDR Text Key178639077
Report Number9610612-2019-00746
Device Sequence Number1
Product Code HTX
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFF803R
Device Catalogue NumberFF803R
Device Lot Number4507865799
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2019
Initial Date Manufacturer Received 10/08/2019
Initial Date FDA Received11/06/2019
Supplement Dates Manufacturer Received11/08/2019
Supplement Dates FDA Received04/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-