The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction tubing leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot h104 was conducted.There were no non-conformance's associated with this lot.This lot met all release requirements.A review of kit lot h104 shows no trends.Trends were reviewed for complaint category, tubing leak.No trends were detected for this complaint category.Photographs were provided by the customer for evaluation.Visual inspection of the photographs confirms a small pool of blood next to the recirculation pump head and a scrape/cut in the recirculation pump loop tubing.The damage to the tubing is at the location where there is an edge on the instrument pump head.The size, shape, and location of the observed damage on the pump loop is consistent with damage that occurs if the pump loop tubing is rubbed against the pump head during installation by the end user.If the pump loop is forcefully rubbed against the pump head during installation of the kit, it could potentially result in scraping and/or cutting of the tubing loop as the pump head rotates.It is unlikely the return pump tubing segment was damaged prior to product release as an in-process leak test is performed on all cellex kits prior to packaging.A material trace of the tubing used to manufacture lot h104 showed no non-conformance.A device history record review did not result in any related non-conformances.This lot passed all lot release testing.The investigation determined that the root cause of the tubing leak was most likely caused by damage that occurred during installation of the pump loop.The cause for the alarm #1: air detected is due to the damaged tubing.No manufacturing related defects were identified through this investigation.No further action is required at this time.This investigation is now complete.Mc: (b)(4).S.D.A.(b)(6) 2019.
|