Model Number NOT APPLICABLE
Device Problems Break (1069); Gas/Air Leak (2946)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/07/2019
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunctions centrifuge bowl leak/break and drive tube leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot h118 was conducted.There were no non-conformance's associated with this lot.This lot met all release requirements.A review of kit lot h118 for the reported issue shows no trends.Trends were reviewed for complaint categories, centrifuge bowl leak/break and drive tube leak/break.No trends were detected for each complaint category.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.Investigation complete.Mc: (b)(4).P.T.(b)(6) 2019.
 
Event Description
The customer contacted mallinckrodt to report they experienced a centrifuge bowl leak/break and drive tube leak/break with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported the centrifuge bowl and drive tube break occurred during the purging air phase of the procedure when 150 ml of whole blood was processed.The customer aborted the ecp treatment and did not return residual blood to the patient.The customer reported that the patient was in stable condition.The customer discarded the kit and will not be returning product for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS, INC.
bedminster NJ 07921
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin, D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
1425 us route 206
bedminster, NJ 07921
MDR Report Key9286742
MDR Text Key207130419
Report Number2523595-2019-00128
Device Sequence Number1
Product Code LNR
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/01/2021
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXECP
Device Lot NumberH118
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/07/2019
Initial Date FDA Received11/06/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age62 YR
-
-