• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD F&P SLEEPSTYLE AUTO CPAP; BZD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD F&P SLEEPSTYLE AUTO CPAP; BZD Back to Search Results
Model Number SPSAAJ
Device Problems Break (1069); Disconnection (1171); Physical Resistance/Sticking (4012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/02/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint f&p sleepstyle series cpap is currently en route to fisher & paykel healthcare in (b)(4) for evaluation.We will provide a follow-up report upon completion of our investigation.
 
Event Description
A distributor in (b)(6) reported that upon removing the power cord from the back of the f&p sleepstyle auto cpap, the inlet pin from the power inlet socket had come apart and was stuck with the power cord.There was no reported patient consequence.
 
Manufacturer Narrative
(b)(4).Method: the complaint f&p sleepstyle series cpap was received at fisher & paykel healthcare in new zealand and was visually inspected and externally.Results: visual inspection confirmed that one pin had detached from the mains inlet socket.Conclusion: the reported incident was traced to an issue in the assembly process of the supplied mains inlet connector component.The supplier of the component was notified and they have made changes to the assembly process.As part of our ongoing product improvement initiatives, we recently implemented a gauge test which identifies and rejects any potentially faulty mains inlet sockets prior to assembly into the sleepstyle.The subject sleepstyle was manufactured prior to implementation of these measures.Our user instructions that accompany the f&p sleepstyle state the following: do not use if the device, power cord or accessories are damaged, deformed, or cracked.Do not pull on the power cord as it may become damaged.Turn the device off at the power supply, then remove the power cord from the rear of the device.
 
Event Description
A distributor in japan reported that upon removing the power cord from the back of the f&p sleepstyle auto cpap, the inlet pin from the power inlet socket had come apart and was stuck with the power cord.There was no reported patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
F&P SLEEPSTYLE AUTO CPAP
Type of Device
BZD
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
MDR Report Key9288710
MDR Text Key178570220
Report Number9611451-2019-01059
Device Sequence Number1
Product Code BZD
UDI-Device Identifier09420012445650
UDI-Public(01)09420012445650(10)2100592135(11)180924
Combination Product (y/n)N
PMA/PMN Number
K173193
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 10/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberSPSAAJ
Device Catalogue NumberSPSAAJ
Device Lot Number2100592135
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2019
Initial Date Manufacturer Received 10/08/2019
Initial Date FDA Received11/06/2019
Supplement Dates Manufacturer Received01/31/2020
Supplement Dates FDA Received02/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-