Catalog Number 688650 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/01/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect device has been returned for evaluation.A follow up will be submitted when the evaluation is complete.
|
|
Event Description
|
The account alleges that during the hemodynamic monitoring of a patient, the pressure monitoring set was found to be leaking fluid at the connections before use.The pm set was exchanged for a new set and the monitoring successfully completed.No patient injury to report.
|
|
Manufacturer Narrative
|
The suspect device has been returned for evaluation.The product was examined visually.The complaint is confirmed.The root cause was attributed to significant force being applied to the device during use.A review of the device history and complaint database could not be performed since the lot number was not provided.
|
|