There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported problem.The plastic insulation of the neutral pin of the plug was melted.According to the technician this was likely caused by a loosening of the wire in the connector.A new mains plug was fitted and the issue solved.Functional verification testing was completed without further issues and the unit was returned to service.
|