• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Melted (1385)
Patient Problem No Patient Involvement (2645)
Event Date 10/11/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported problem.The plastic insulation of the neutral pin of the plug was melted.According to the technician this was likely caused by a loosening of the wire in the connector.A new mains plug was fitted and the issue solved.Functional verification testing was completed without further issues and the unit was returned to service.
 
Event Description
Livanova (b)(4) received a report of a melted plug of a heater-cooler system 3t.The user also reported that also the theatre mains wall socket was slightly damaged.There was no patient involvement.
 
Manufacturer Narrative
H.10: pictures of the plug have been provided.The analysis of the photos revealed a wrongly de-isolated wiring and an insufficient fixing of the wiring with the screw leading to high temperature and melting of the plug.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich
MDR Report Key9294363
MDR Text Key168711579
Report Number9611109-2019-00879
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/11/2019
Initial Date FDA Received11/08/2019
Supplement Dates Manufacturer Received10/25/2019
Supplement Dates FDA Received11/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-