• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.O.M. WORLD OF MEDICINE GMBH AQUILEX FLUID CONTROL SYSTEM; HYSTEROSCOPY PUMP AND SCALE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.O.M. WORLD OF MEDICINE GMBH AQUILEX FLUID CONTROL SYSTEM; HYSTEROSCOPY PUMP AND SCALE Back to Search Results
Model Number AQL-100CS
Device Problem Defective Device (2588)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
We have been informed of the following event: "tech support was contacted by (b)(6) territory manager, to report aquilex system not working consistently with accuracy.Pump serial (b)(4) reported in case (b)(4).Cart serial (b)(4) reported in case (b)(4).Contact at facility is (b)(6).(b)(6) noted customer has reported a deficit that is 1500 higher than is should be at times as well as occasionally registering a negative deficit.Mike assisted customer in running the fluid deficit testing and it did not come out within the permitted limits.Unit was pulled from service.They checked that all tubing was connected correctly, nothing was hanging on the white canister scale holder, no obvious reason why this is not working correctly.No patient injury reported.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUILEX FLUID CONTROL SYSTEM
Type of Device
HYSTEROSCOPY PUMP AND SCALE
Manufacturer (Section D)
W.O.M. WORLD OF MEDICINE GMBH
salzufer 8
berlin, 10587
GM  10587
Manufacturer (Section G)
W.O.M. WORLD OF MEDICINE GMBH
alte poststrasse 11
ludwigsstadt, 96337
GM   96337
Manufacturer Contact
lucia puettmann
salzufer 8
berlin, 10587
GM   10587
MDR Report Key9295100
MDR Text Key196470243
Report Number3002914049-2019-00016
Device Sequence Number1
Product Code HIG
UDI-Device Identifier04056702002145
UDI-Public04056702002145
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112642
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAQL-100CS
Device Catalogue NumberAQL-100CS
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/11/2019
Initial Date FDA Received11/08/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/22/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-