MAUDE Adverse Event Report: EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number VO2346TTPL
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Output below Specifications (3004)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 10/17/2019
Event Type  malfunction  
Manufacturer Narrative
One dpt was returned for examination.The reported event of pressure measurement issue was not confirmed.The dpt zeroed and sensed pressure accurately on a pressure monitor.The pressure did not show any drift during output drift testing and met specification.Electrical testing showed that the dpt electronic components were intact because both input and output impedance were within specifications.Zero-offset also met specification.No visible defect was found from the dpt cable connector.A supplemental report will be forthcoming with the device history results when received.Poor dynamic response can be caused by air bubbles, clotting, and excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.The assembly may be tested for dynamic response by observing the pressure waveform on an oscilloscope or monitor.Bedside determination of the dynamic response of the catheter, monitor, kit and transducer system is done after the system is flushed, attached to the patient, zeroed and calibrated.A square-wave test may be performed by pulling the snap tab device and releasing quickly.Pressure readings can change quickly and dramatically because of loss of proper calibration, loose connection, or air in the system.Pressure readings should correlate with the patient¿s clinical manifestations.It is not known if user or procedural factors may have contributed to the stated event.In this event, there was no patient compromise noted.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that a value was lower than the expected value during use of this disposable pressure transducer (dpt).Which value had defect and the expected value are unknown.It was able to zero.The device was exchanged, and the problem was solved.The sample of tracing was not available.Patient demographic information requested but unavailable.There were no patient complications reported.
 
Manufacturer Narrative
A review of the manufacturing records indicated that the product met specifications upon release.
 
Manufacturer Narrative
Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE MONITORING KIT
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR 
MDR Report Key9300413
MDR Text Key195970782
Report Number2015691-2019-04170
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 10/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/15/2023
Device Model NumberVO2346TTPL
Device Lot NumberLJ0838MT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2019
Initial Date Manufacturer Received 10/17/2019
Initial Date FDA Received11/08/2019
Supplement Dates Manufacturer Received11/29/2019
07/23/2020
Supplement Dates FDA Received12/12/2019
01/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-