• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA DRAGONFLY¿ OPTIS¿ KIT BOX IMAGING CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA DRAGONFLY¿ OPTIS¿ KIT BOX IMAGING CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number C408646
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thrombosis (2100)
Event Date 10/29/2019
Event Type  Injury  
Event Description
When the catheter was pulled out of the patient there was a blood clot present on the catheter, and was unable to be purged from the catheter.Another catheter was used to continue and complete the procedure.
 
Manufacturer Narrative
One dragonfly optis catheter was returned to the manufacturer for analysis.The results of the investigation were inconclusive due to the condition of the returned device.There was dried contrast and blood-like substance inside the catheter lumen which was unable to be purged.The catheter was submerged in an ultrasonic water bath to remove the dried contrast; however, the dried contrast was still unable to be removed and purged, which prevented further functional testing.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the received condition of the device and the information provided, the cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRAGONFLY¿ OPTIS¿ KIT BOX IMAGING CATHETER
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key9309736
MDR Text Key168228067
Report Number3008452825-2019-00570
Device Sequence Number1
Product Code DQO
UDI-Device Identifier00183739000654
UDI-Public00183739000654
Combination Product (y/n)N
PMA/PMN Number
K141769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2021
Device Model NumberC408646
Device Lot Number7094641
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/29/2019
Initial Date FDA Received11/12/2019
Supplement Dates Manufacturer Received12/26/2019
Supplement Dates FDA Received01/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-