Catalog Number 8065751763
Device Problems Thermal Decomposition of Device (1071); Smoking (1585)
Patient Problem Injury (2348)
Event Date 10/08/2019
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that fluid management system and handpiece was tested and successfully passed.The surgeon indicated that during sculpting the phaco burn, smoky liquid and damage occurred during procedure in main incision.The nurse did not feel any resistance in the handpiece during flushing.Surgeon continued the surgery with the same set of instruments and finished the surgery.The cornea took time to stabilize.Patient did not come back to follow-up after 2 weeks.
 
Manufacturer Narrative
The lot complaint history was reviewed; this is the first complaint for the finish goods lot and first for this issue for this lot.The device history record shows the product was released per specifications.No sample has been returned for evaluation; therefore, the condition of the product could not be verified.The root cause of the customer's complaint could not be established as a sample has not been received and the condition of the product could not be verified.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.The root cause for this complaint is not known, therefore, specific action with regards to this complaint cannot be taken.After final assembly each cassette is verified that all required tests have been performed and all acceptance criteria are met prior to release.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Consumables manufacturing management has been made aware of this complaint through the consumer affairs review meeting.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key9314849
MDR Text Key166143633
Report Number2028159-2019-02059
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 02/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751763
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/04/2019
Initial Date FDA Received11/13/2019
Supplement Dates Manufacturer Received02/03/2020
Supplement Dates FDA Received02/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CENTURION SURGICAL PROCEDURE PAK; HANDPIECE; CENTURION SURGICAL PROCEDURE PAK; HANDPIECE
Patient Outcome(s) Other;
-
-