Returned product consisted of an ffr comet pressure wire connected to the occ cable.The tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The shaft showed two kinks located 151.5cm from the tip and 176cm from the tip.There was peeled coating located at the 176cm location.The occ handle was connected to the ffr link for signal verification.The signal was not present as designed.Device analysis was conducted by inspecting the proximal end of the wire for any damage to the fiber optic.No damage was noticed.The sensor was inspected by viewing the sensor port to verify that the sensor was in the correct location.This sensor looked to be in the correct location.The wire was gently shaken to see if the sensor would move within the sensor housing and the sensor did move which verifies the sensor was detached from the fiber optic.The occ handle cap was loosened to remove the wire.There was no issue with removing the wire.The sensor port showed no residue of body fluids.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
Reportable based on analysis completed on 30oct2019.It was reported that no signal was present.A comet pressure guidewire was being used; however, no signal was recognized by the wifi box.The procedure was successfully completed with a different comet pressure guidewire without issue or patient injury.However, returned device analysis revealed peeled coating.
|