Catalog Number S-65-100-120-P6
Device Problems Entrapment of Device (1212); Material Separation (1562)
Patient Problem Vessel Or Plaque, Device Embedded In (1204)
Event Date 10/30/2019
Event Type  Injury  
Manufacturer Narrative
Exemption number e2019001.The customer reported the delivery system was discarded, but the stent remains in the patient.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was to treat a superficial femoral artery (sfa).After introducing a 6.5x100mm supera self-expanding stent delivery system into the anatomy, the tip of the stent broke off while attempting to deploy.The nose cone got caught at the end of the stent, however the stent was able to be deployed successfully.The physician did not have a snare device to remove the tip, so a second stent was deployed over the tip with a good result.The tip still remains in the anatomy.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents and/or complaints.The investigation determined that the reported difficulties were due to procedural circumstances.Based on the information provided, the tip detachment occurred due to the tip catching on the distal end of the stent during deployment.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA PERIPHERAL STENT SYSTEM
Type of Device
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9315282
MDR Text Key166167715
Report Number2024168-2019-13495
Device Sequence Number1
Product Code NIP
UDI-Device Identifier08717648226281
UDI-Public08717648226281
Combination Product (y/n)N
PMA/PMN Number
P120020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2020
Device Catalogue NumberS-65-100-120-P6
Device Lot Number8121461
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/30/2019
Initial Date FDA Received11/13/2019
Supplement Dates Manufacturer Received12/27/2019
Supplement Dates FDA Received01/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight100
-
-