Model Number 27-80-20
Device Problem Failure to Auto Stop (2938)
Patient Problem No Patient Involvement (2645)
Event Date 10/17/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 cardioplegia sensor module.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and the issue could not be duplicated.The device will be replaced with a loaner.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 cardioplegia sensor module probably failed during setup.The used noticed that the cardioplegia pump was off thus, it was turned on in the middle.The delivery of cardioplegia solution was started but the cardioplegia pump did not stop when the set amount of cardioplegia solution was delivered.Moreover, the total volume count did not appear on the display.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 CARDIOPLEGIA SENSOR MODULE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9316353
MDR Text Key196428682
Report Number9611109-2019-00895
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 11/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number27-80-20
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/17/2019
Initial Date FDA Received11/13/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-