PHILIPS MEDICAL SYSTEMS PIIC IX HARDWARE; CENTRAL STATION MONITOR Back to Search Results
Model Number 866424
Device Problems Device Alarm System (1012); Component Missing (2306)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/05/2019
Event Type  Injury  
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Event Description
The customer reported that retrospective data was missing and the patient was defibrillated.A patient was admitted since (b)(6) 2019, but the data is only available since the night from (b)(6) 2019 to (b)(6) 2019 after 4:50am in the control center.They defibrillated in the meantime.The patient was not admitted and was in bed 2t.The inability to access retrospective data would not prevent the primary monitoring device from continuing to provide real-time monitoring and alarming; however, since treatment was required for the patient, and it is unknown if the device was a factor in the patient event.
 
Manufacturer Narrative
A quote was provided to the customer for assistance with the retrospective data, however, no further response has been received from the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIIC IX HARDWARE
Type of Device
CENTRAL STATION MONITOR
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
betty harris
3000 minuteman road
andover, MA 01810
9786871501
MDR Report Key9321994
MDR Text Key166362153
Report Number1218950-2019-08647
Device Sequence Number1
Product Code MHX
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K102495
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866424
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/05/2019
Initial Date FDA Received11/14/2019
Supplement Dates Manufacturer Received11/05/2019
Supplement Dates FDA Received12/18/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/17/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-