BOSTON SCIENTIFIC CORPORATION ULTRAFLEX ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL Back to Search Results
Device Problems Use of Device Problem (1670); Component Missing (2306); Obstruction of Flow (2423)
Patient Problems Discomfort (2330); No Code Available (3191)
Event Date 10/21/2019
Event Type  Injury  
Manufacturer Narrative
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.The complainant was unable to provide the suspect device upn and lot number.Therefore, the manufacture and expiration dates are unknown.According to the complainant the stent was implanted at (b)(6).(b)(4).The complainant indicated that the suspect device remains implanted and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2019 that an ultraflex esophageal uncovered stent has been implanted to treat a leak in the esophagus during a gastroscopy with stent placement procedure performed approximately 3 months earlier.According to the complainant, on (b)(6) 2019, three (3) months post stent placement, the patient presented with discomfort.A gastroscopy procedure was performed and the stent was found to have tissue ingrowth along its length.Reportedly, the physician planned to bring the patient back to remove the stent on (b)(6) 2019.Despite efforts, boston scientific has been unable to confirm if the stent was successfully removed.The patient's condition at the conclusion of the procedure was reported to be stable and the event was reported to be resolved.Note: according to the complainant, the ultraflex esophageal uncovered distal release stent was placed to treat leak.However, per ultraflex esophageal ng stent system directions for use, the stent is indicated for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors only.
 
Manufacturer Narrative
Blocks b5, d7, h6 and h10 have been updated with additional information received on november 14, 2019, december 01, 2019 and december 05, 2019.Block a2: patient's exact age is unknown; however it was reported that the patient was over the age of 18.Block d4, h4: the complainant was unable to provide the suspect device upn and lot number.Therefore, the manufacture and expiration dates are unknown.Block e1: according to the complainant the stent was implanted at cairns base hospital by dr.(b)(6).Block h6: problem code 2423 captures the reportable event of stent blocked/ occluded.Problem code 2306 captures the reportable event of stent cover missing.Patient code 3191 is being used to capture the additional intervention to remove the stent.Block h10: according to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation on october 22, 2019 that an ultraflex esophageal uncovered stent has been implanted to treat a leak in the esophagus during a gastroscopy with stent placement procedure performed approximately 3 months earlier.According to the complainant, on october 21, 2019, three (3) months post stent placement, the patient presented with discomfort.A gastroscopy procedure was performed and the stent was found to have tissue ingrowth along its length.Reportedly, the physician planned to bring the patient back to remove the stent on (b)(6) 2019.Despite efforts, boston scientific has been unable to confirm if the stent was successfully removed.The patient's condition at the conclusion of the procedure was reported to be stable and the event was reported to be resolved.Note: according to the complainant, the ultraflex esophageal uncovered distal release stent was placed to treat leak.However, per ultraflex esophageal ng stent system directions for use, the stent is indicated for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors only.Additional information received on november 14, 2019.According to the complainant, during the scheduled stent removal procedure on (b)(6) 2019, the stent was found to still have tissue in growth so another stent was implanted through it.Additional information received on december 01, 2019 and december 05, 2019.According to the complainant, the stent was successfully removed from the patient on (b)(6) 2019 using an overtube and rat tooth grasping forceps.Additionally, it was reported that the stent was originally placed to treat a post-surgical gastric sleeve leak.The original implanting physician clarified that the stent was not an uncovered stent at implantation.The stent was believed to be a covered stent when originally placed.The lack of a stent covering was not noticed during the index procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX ESOPHAGEAL
Type of Device
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key9323496
MDR Text Key166998626
Report Number3005099803-2019-05567
Device Sequence Number1
Product Code ESW
Combination Product (y/n)N
PMA/PMN Number
K955347
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/22/2019
Initial Date FDA Received11/14/2019
Supplement Dates Manufacturer Received11/14/2019
Supplement Dates FDA Received12/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-