This follow-up report is being submitted to relay additional information.The complaint is confirmed.The complaint was opened because the facility reported the instrument tip broke off.The elevator #301 (09-0257, lot# 040617c17 ) was returned for investigation.Visual evaluation showed signs of use as there was minor scratching on the body of the elevator and the tip had fractured off.The dhr was reviewed and there were no non-conformances found.There are no indications of a manufacturing defects and there are no updates to the risk documents needed.Complaint history for (b)(4), lot# 040617c17 was reviewed and there is a complaint rate of 1.6% which is no greater than the occurrence listed in the application fmea.The most likely underlying cause of the complaint is that excessive force was used, beyond what the instrument was designed to encounter.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.The following fields were updated: b4 date of this report.B5 describe event or problem.D4 lot number.D10 device availability.G4 date received by manufacturer.G7 type of report.H2 follow up type.H3 device evaluated by manufacturer.H4 device manufacturer date.H6 method code.H6 results code.H6 conclusions code.H10 additional narratives/data.
|