• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MERLIN PROGRAMMER; PACEMAKER PROGRAMMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MERLIN PROGRAMMER; PACEMAKER PROGRAMMER Back to Search Results
Model Number 3650
Device Problems Failure to Power Up (1476); Sparking (2595)
Patient Problem No Patient Involvement (2645)
Event Date 11/06/2019
Event Type  malfunction  
Manufacturer Narrative
The reported event of failure to power up was confirmed by analysis.Final analysis found a failed inverter board to be the cause of the failure to power up.During analysis, a failure event was observed which was unrelated to the reported event.Final analysis found the power switch made a crackling sound and emitted sparks when turned on.Testing confirmed a failed power switch was the cause of the sparks.The power supply and inverter board were replaced to resolve the event.
 
Event Description
The programmer was returned to the manufacturer when it would not power up.This report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Type of Device
PACEMAKER PROGRAMMER
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key9331476
MDR Text Key166707227
Report Number2017865-2019-15613
Device Sequence Number1
Product Code KRG
UDI-Device Identifier05414734500661
UDI-Public05414734500661
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial
Report Date 11/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3650
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/06/2019
Initial Date FDA Received11/15/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/12/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-