Catalog Number 2400L
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 10/25/2019
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation.During visual inspection, the device was found to have all four (4) feet bent and very difficult to fit the load cell into the base plate.The reported condition was verified.The cause of the condition could not be determined.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the legs of the load cell on the em 2400 were observed "bent" and "won't fit into the base unit".There was no patient involved.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EM2400 LOAD CELL MODULE
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
ECM - ENGLEWOOD
14445 grasslands dr
englewood CO 80112
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9341532
MDR Text Key167557437
Report Number1416980-2019-06417
Device Sequence Number1
Product Code NEP
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 11/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2400L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/25/2019
Initial Date FDA Received11/19/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N