A customer reported that the device did not power on while in use on a patient and that another device was used.The device was reported to be in use on a patient, but no adverse event to patient or user was reported.A philips field service engineer (fse) evaluated the device and discovered a faulty processor pca.The device passed all performance tests.No ecg strips or case events files were provided to philips for evaluation.The information gathered for this report does not provide evidence of a systemic, design, or labeling problem.No further investigation or action is warranted.
|