Catalog Number 03666794001
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Patient Involvement (2645)
Event Date 10/30/2019
Event Type  malfunction  
Manufacturer Narrative
The customer¿s meter was requested for return.The product has not been received at this time.If the product is returned in the future, a follow up report will be submitted.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Event Description
The initial reporter complained of a display issue with coaguchek xs meter serial number (b)(4).The display issue complained about could cause results to be misinterpreted.The customer stated that they performed a full screen display on their meter and that the top halves of the 88.8 in the results area were missing.There was no misinterpretation of results.
 
Manufacturer Narrative
The customer's meter was received for investigation.The meter¿s circuit board was tested for damage or contamination and it was found the printed circuit board (pcb) was contaminated.The root cause is determined to be contamination of the contacts due to improper handling or maintenance by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key9353526
MDR Text Key219127436
Report Number1823260-2019-04156
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2019
Initial Date Manufacturer Received 10/31/2019
Initial Date FDA Received11/21/2019
Supplement Dates Manufacturer Received10/31/2019
Supplement Dates FDA Received12/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-