Catalog Number 03666794001 |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/30/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer¿s meter was requested for return.The product has not been received at this time.If the product is returned in the future, a follow up report will be submitted.The investigation did not identify a product problem.The cause of the event could not be determined.
|
|
Event Description
|
The initial reporter complained of a display issue with coaguchek xs meter serial number (b)(4).The display issue complained about could cause results to be misinterpreted.The customer stated that they performed a full screen display on their meter and that the top halves of the 88.8 in the results area were missing.There was no misinterpretation of results.
|
|
Manufacturer Narrative
|
The customer's meter was received for investigation.The meter¿s circuit board was tested for damage or contamination and it was found the printed circuit board (pcb) was contaminated.The root cause is determined to be contamination of the contacts due to improper handling or maintenance by the customer.
|
|