LIMACORPORATE SPA SMR CTA HUMERAL HEAD Ø42 MM; SMR CTA HUMERAL HEAD Ø42 MM (HSD-KWS) Back to Search Results
Model Number 1323.09.420
Device Problem Material Erosion (1214)
Patient Problem Arthralgia (2355)
Event Date 11/13/2019
Event Type  Injury  
Manufacturer Narrative
By checking the dhr of the lot #: 0901703 ster.1500086 no anomaly was found.This is the first and only complaint received on this lot#.We will submit a final mdr once the investigation will be concluded.
 
Event Description
Revision surgery due to painful joint occurred on (b)(6) 2019.Previous surgery was performed on (b)(6) 2017.During revision surgery, smr cta humeral head code 1323.09.420 lot #: 0901703 was explanted.Revision stem was left in situ.Patient has had a hemiarthroplasty with a cta head since the last revision.According to the information reported, the cta head was riding high on the acromian, eroding the glenoid and causing pain.No future revision is planned unless pain does not subside.Event occurred in (b)(6).
 
Event Description
Revision surgery due to painful joint occurred on (b)(6) 2019.Previous surgery was performed on (b)(6) 2017.During revision surgery, smr cta humeral head code 1323.09.420 lot # 0901703 was explanted.Revision stem was left in situ.According to the information reported, the cta head was riding high on the acromian, eroding the glenoid and causing pain.No future revision is planned unless pain does not subside.Event occurred in australia.
 
Manufacturer Narrative
Check of the dhr: by checking the manufacturing chart of the lot # 0901703 ster.1500086 no pre-existing anomaly was found on 20 smr cta humeral heads manufactured.This is the first and only complaint received on this lot#.Ct-slices analysis: we received pre-revision ct-slices (exact date unknown) and sent them to our medical consultant for analysis.Following, his comment is reported: "as the report suggests the cta head has migrated superiorly out of the glenoid and into the supraspinatus fossa.The one redeeming feature is the head is still contained within the coracoacromial arch.There has been erosion of the acromion and outer end of clavicle and apparent extreme bone loss from the glenoid but these are ct slices and do not likely truely represent the total glenoid.The humeral component appears well fixed with no obvious features of loosening.We know from experience that a cta head is almost never associated with a good patient outcome so it is not surprising the patient has pain.That said pain in revision situations should always raise suspicion for infection.In summary while no doubt the patient is unhappy this is the expected outcome of a cta head.I do not see any technical aspects of the surgery that would cause concern." according to the medical consultant, ct-slices provided by the complaint source do not allow a complete analysis of the case.However, he confirmed that cta head migrated superiorly into the supraspinatus fossa, but it was still within the coracoacromial arch.It was also confirmed that there has been erosion of the acromion and outer end of clavicle and apparent extreme bone loss from the glenoid.According to the medical consultant, there were no aspects related to the surgery or to the product itself related to the negative abovementioned outcomes.A further possible cause for pain could have been infection, but the complaint source did not report any swab analysis.Considering that: all the components manufactured with lot #0901703 were up to specifications and no further complaints were received on this lot# the medical consultan did not find any anomaly in the surgery that could have been responsible for failure of the prosthesis we can conclude that the event is due to patient clinical condition.Event not product related.Pms data according to our pms data, revision rate of cta heads (family of code 1323.09.Xxx - 1323.15.Xxx) due to pain is 0.08%.None of these event was judged as product related.No corrective/preventive action implemented for this specific case.Limacorporate will keep the market monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMR CTA HUMERAL HEAD Ø42 MM
Type of Device
SMR CTA HUMERAL HEAD Ø42 MM (HSD-KWS)
Manufacturer (Section D)
LIMACORPORATE SPA
via nazionale 52
villanova di san daniele, udine 33038
IT  33038
MDR Report Key9353660
MDR Text Key167351362
Report Number3008021110-2019-00130
Device Sequence Number1
Product Code HSD
Combination Product (y/n)N
PMA/PMN Number
K110847
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 05/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Model Number1323.09.420
Device Lot Number0901703
Initial Date Manufacturer Received 11/14/2019
Initial Date FDA Received11/21/2019
Supplement Dates Manufacturer Received11/14/2019
Supplement Dates FDA Received05/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-