Patient underwent device explant of the lead and generator due to no longer desiring the device.Lead product analysis was completed and revealed breaches in the inner tubing insulation.Generator product analysis was completed.In the product analysis lab, the device output signal was monitored for more than 24-hrs, while the generator was placed in a simulated body temperature environment.Results showed no signs of variation in the pulse generator¿s output signal and demonstrated that the device provided the expected level of output current for the entire monitoring period.The pulse generator diagnostics were as expected for the programmed parameters.There were no performance of any other type of adverse events found with the pulse generator.Lead product analysis was completed.The inner silicone tubing was noted to be abraded open, exposing a portion of the negative coil.Also, scanning electron microscopy images of the negative coil suggest the strands were exposed to some type of electro-cautery tool as indicated by their melted appearance.Note that since a significant portion of the lead (including the electrode array) was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.Other than the above mentioned observations and typical wear and explant related observations, no other anomalies were identified in the returned lead portions.No other relevant information has been received to date.
|