• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVACARE FLORIDA SC900DLX; BED, THERAPEUTIC, AC-POWERED, ADJUSTABLE HOME-USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVACARE FLORIDA SC900DLX; BED, THERAPEUTIC, AC-POWERED, ADJUSTABLE HOME-USE Back to Search Results
Model Number NA:IHSC900DLX
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The facility maintenance representative advised that the issue was discovered when they were performing an inspection of all rails at the facility.It was observed that a total of 74 pairs of rails had at least one missing end cap.It is unknown at what point they went missing.The maintenance rep advised that the facility has a rail maintenance program to inspect the rails, but he was not aware of the last time it was completed.He stated that the residents are still using the rails with the missing end caps.He was aware that they shouldn¿t be using them, but he indicated that they had no choice.No injury or damage was reported as a result of the missing end caps.The date code of the rails is unknown; however, the maintenance rep advised that they are approximately 3 years old.He was unable to provide any further information regarding the rails.The underlying cause of the end caps falling out is undetermined.This issue has been previously investigated.The following statement was added to the ihrailae-dlx installation user manual: "to avoid death, injury or damage due to improper maintenance or inspection.Always maintain and inspect equipment per the instructions in this manual.Contact a qualified technician or invacare if any of the following issues are present: loose or missing parts such as end caps, knobs, bolts, screws etc., should be secured or replaced.Sharp edges or surfaces should be corrected or replaced." a design change has since been made to these assist rails, as of 10/26/2017.The plastic end cap has been replaced with a permanent steel cover, which is brazed onto the tube ends and buffed to create a smooth edge.Should additional information become available, a supplemental record will be filed.
 
Event Description
A maintenance representative from a facility reported that the end caps are falling out of the ihrailae-dlx assist rails, which are being used on ihsc900dlx beds reported.A total of 74 records have been entered into the system regarding this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SC900DLX
Type of Device
BED, THERAPEUTIC, AC-POWERED, ADJUSTABLE HOME-USE
Manufacturer (Section D)
INVACARE FLORIDA
2101 east lake mary blvd
sanford FL 32773
Manufacturer (Section G)
INVACARE FLORIDA
2101 east lake mary blvd
sanford FL 32773
Manufacturer Contact
jason fiest
one invacare way
elyria, OH 44036
8003336900
MDR Report Key9364401
MDR Text Key219594533
Report Number1031452-2019-00105
Device Sequence Number1
Product Code LLI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberNA:IHSC900DLX
Device Catalogue NumberIHSC900DLX
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/31/2019
Initial Date FDA Received11/22/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-