DEPUY ORTHOPAEDICS INC US UNK FIXATION PIN Back to Search Results
Catalog Number UNK FIXATION PIN
Device Problems Break (1069); Mechanical Jam (2983); Physical Resistance/Sticking (4012)
Patient Problem Not Applicable (3189)
Event Date 10/29/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the a headed pin got stuck in a tc3 cut block and when trying to remove it, the head of the pin was broken off and became stuck in the pin driver.A surgical delay of 5 minutes.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: product complaint # (b)(4).Investigation summary examination of the returned device confirmed the reported event.The investigation attributed the root cause to unintended misuse/misapplication of the device and did not indicate that that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot null.Device history batch null.Device history review null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: product complaint # (b)(4).Investigation summary examination of the returned device confirmed the reported event.The investigation attributed the root cause to unintended misuse/misapplication of the device and did not indicate that that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot null.Device history batch null.Device history review null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK FIXATION PIN
Type of Device
FIXATION PIN
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9364506
MDR Text Key187456594
Report Number1818910-2019-117652
Device Sequence Number1
Product Code HTY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 10/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK FIXATION PIN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2019
Initial Date Manufacturer Received 10/29/2019
Initial Date FDA Received11/22/2019
Supplement Dates Manufacturer Received12/19/2019
12/19/2019
Supplement Dates FDA Received12/20/2019
12/20/2019
Patient Sequence Number1
Patient Age83 YR
-
-