Model Number BI70000028230
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 11/14/2019
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.It was determined the system would not boot up.The cmos battery was determined to be faulty.The cmos battery was replaced.The system then passed the system checkout and was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging system being used outside of a procedure.There was no patient involvement.The imaging system would not get passed the initializing stage.
 
Manufacturer Narrative
Continuation of concomitant medical products: section 'suspect device' information references the main component of the system.Other relevant device(s) are: product id: bi71000165, serial/lot : na.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM 3RD EDITION
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key9365859
MDR Text Key179455269
Report Number3004785967-2019-02042
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K092564
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberBI70000028230
Device Catalogue NumberBI70000028230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2019
Initial Date FDA Received11/22/2019
Supplement Dates Manufacturer Received11/24/2019
Supplement Dates FDA Received12/19/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-