• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK ATLAS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK ATLAS Back to Search Results
Catalog Number 10309477
Device Problem False Negative Result (1225)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/05/2019
Event Type  malfunction  
Manufacturer Narrative
Siemens field service engineer went onsite.The probe tip was bent, the pipette movement was binding and the sg well was dirty.The engineer replaced the probe, cleaned and lubricated the pipette and cleaned the sg well.The calibration was successful.The system was fully functional on departure.
 
Event Description
The customer reported false negative urine blood results on two patients on the clinitek atlas when compared to another siemens analyzer and confirmed microscopically.There was no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK ATLAS
Type of Device
CLINITEK ATLAS
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SPARTAN MEDICAL
22740 lunn road
cleveland OH 44149
Manufacturer Contact
felix akinrinola
2 edgewater drive
norwood, MA 02062
6107052212
MDR Report Key9370370
MDR Text Key169087099
Report Number3002637618-2019-00124
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K932674
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10309477
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/05/2019
Initial Date FDA Received11/25/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-