|
Model Number 8900 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/18/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluated by mfr: returned product consisted of an ffr comet pressure wire connected to the occ cable.The tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The shaft showed 1 kink located 54cm from the tip.The tip showed kinks/bend damage.There was peeled coating at the 54cm location.The occ handle was connected to the ffr link for signal verification.The signal was present as designed.The occ handle cap was loosened to remove the wire.There was no issue with removing the wire.The sensor housing showed residue of body fluids.The wire was inserted into the pressure chamber test equipment and the pressure was increased to verify the sensor was indeed reacting to the pressure increases and decreases.The pressure sensor functioned as designed.The coefficient was confirmed to be in specification.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
|
Event Description
|
It was reported that a pressure signal was not recognized.During preparation of a ffr procedure and outside the patient body, a comet wire was connected to an ffr link, but the comet was not recognized by the polaris.Hub reconnection was performed, but the issue was not resolved.The procedure was completed with a non bsc device.No patient complications resulted from this event.However, device analysis revealed peeled coating.
|
|
Search Alerts/Recalls
|
|
|