Model Number R SERIES
Device Problem Unable to Obtain Readings (1516)
Patient Problem Death (1802)
Event Date 10/01/2019
Event Type  Death  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device was unable to obtain an ecg signal.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
 
Manufacturer Narrative
The customer's report was observed and attributed to a faulty integrated circuit on the analog board.The device was recertified and returned to the customer.Please refer to the medwatch report 1220908-2019-03192 for a medwatch report inadvertently filed for this same patient event.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key9376905
MDR Text Key168034213
Report Number1220908-2019-03653
Device Sequence Number1
Product Code DRT
UDI-Device Identifier00847946002459
UDI-Public00847946002459
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060559/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/07/2019
Initial Date FDA Received11/26/2019
Supplement Dates Manufacturer Received11/07/2019
Supplement Dates FDA Received12/23/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
-
-