Model Number R SERIES |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
Death (1802)
|
Event Date 10/01/2019 |
Event Type
Death
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device was unable to obtain an ecg signal.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
|
|
Manufacturer Narrative
|
The customer's report was observed and attributed to a faulty integrated circuit on the analog board.The device was recertified and returned to the customer.Please refer to the medwatch report 1220908-2019-03192 for a medwatch report inadvertently filed for this same patient event.Analysis of reports of this type has not identified an increase in trend.
|
|