Device returned to manufacturer: the comet was returned and analysis was completed.The tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The shaft showed two kinks.The first kink was located 38cm from the tip.The second kink was located 177cm from the tip.There was tip damage noticed.There was peeled coating at the 177cm location.The wire was connected to the returned occ handle.The occ handle was connected to the ffr link for signal verification.The signal was not present as designed.The occ handle cap was loosened to remove the wire.There was no issue with removing the wire.A known good occ handle was used to verify cable operation was correct with the returned occ.The test occ cable and the connected returned wire functioned as designed.The returned cable was dissected to check for damage.It was noticed that the fep disk was damaged where the wire seats against the disk.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|