• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MILACA, INC. DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MILACA, INC. DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 11/06/2019
Event Type  malfunction  
Manufacturer Narrative
Product analysis: analysis confirmed the customer comment that the output connector assembly was broken.It was noted that the upper case was broken, encoder assembly and four knobs were contaminated, lower case was broken, display wires were pinched, wire insulation was exposed, and six case screws were contaminated.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that the epg was returned for repair.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the atrial receptible on the external pulse generator (epg) was broken.The status of the epg is unknown.There was no patient involvement.It was further reported that the epg was returned for repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
MEDTRONIC MILACA, INC.
900 sixth avenue ne
milaca MN 56353 3728
Manufacturer (Section G)
MEDTRONIC MILACA, INC.
900 sixth avenue ne
milaca MN 56353 3728
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9379439
MDR Text Key170559379
Report Number2183613-2019-00204
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/06/2019
Initial Date FDA Received11/26/2019
Supplement Dates Manufacturer Received12/03/2019
Supplement Dates FDA Received12/12/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/04/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-