Model Number 8900
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/10/2019
Event Type  malfunction  
Manufacturer Narrative
Returned product consisted of an ffr comet pressure wire connected to the occ cable.The tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The shaft showed 2 kinks located 157cm and 175cm from the tip.The tip showed damage in the form of bends.There was some peeled coating at the kinked locations.The occ handle was connected to the ffr link for signal verification.The signal was present as designed.The occ handle cap was loosened to remove the wire.There was no issue with removing the wire.The sensor housing showed no residue of body fluids.The wire was inserted into the pressure chamber test equipment and the pressure was increased to verify the sensor was indeed reacting to the pressure increases and decreases.The pressure sensor functioned as designed.The coefficient was confirmed to be in specification.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.Because there was no evidence of any product quality deficiencies, it was considered likely that the kinks, tip damage and peeled coating were attributable to handling.
 
Event Description
Reportable based on analysis completed on 08 nov 2019.It was reported that recognition failure occurred.A coronary angiogram was being performed on a mildly calcified and moderately tortuous lesion in a coronary artery.While using the comet pressure guidewire, recognition failure occurred during prep.The procedure was successfully completed with a different device without issue or patient injury.However, returned device analysis revealed peeled coating.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMET
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9380354
MDR Text Key173946270
Report Number2134265-2019-14333
Device Sequence Number1
Product Code DXO
UDI-Device Identifier08714729904403
UDI-Public08714729904403
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K151610
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 11/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/08/2021
Device Model Number8900
Device Catalogue Number8900
Device Lot Number0023764281
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/08/2019
Initial Date FDA Received11/26/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/09/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-