Model Number 92215
Device Problem False Positive Result (1227)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/04/2019
Event Type  malfunction  
Manufacturer Narrative
Devices not returned.Product discarded/expended by customer.Investigation conclusion: retained devices from the reported lot number were tested with hcg-negative clinical serum samples.The results were read at 5 minutes and all devices yielded the expected negative results.No false positive results were observed during in-house testing.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review did not uncover any relevant non-conformances and found that the lot met quality control specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.Case details indicate that the concentration of hcg present in the sample was 8 miu/ml.When testing samples near the threshold of 10 miu/ml some positive results may occur.Per the package insert, this test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.
 
Event Description
(b)(6) 2019: a (b)(6) female patient visited the facility due to pelvic pain.The patient's serum specimen was collected at 9:32 am and tested on the alere hcg combo cassette.A positive result was obtained at the 5 minute read time.No treatment was provided based on false positive result.The same serum specimen was retested on the beckman access 2 and a confirmatory hcg quant of 7 miu/ml was obtained.(b)(6) 2019: an additional confirmatory quant test was performed on the beckman access 2 and a result of 6 miu/ml was obtained.Although further information was requested, no further information was provided.No adverse patient outcomes were reported.Troubleshooting occurred with a discussion about possible causes for unexpected results focusing on proper sampling technique, storage conditions and patient specific factors per the package insert (pi)-no deviations noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALERE HCG COMBO CASSETTE (20/10MIU/ML)
Type of Device
PREGNANCY TEST
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9975 summers ridge rd.
san diego CA 92121
Manufacturer Contact
monika burrell
9975 summers ridge rd.
san diego, CA 92121
8588052506
MDR Report Key9387148
MDR Text Key188863072
Report Number2027969-2019-00593
Device Sequence Number1
Product Code JHI
UDI-Public(01)2088538002026817)210228(10)HCG9030087
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Model Number92215
Device Catalogue NumberB1077-23
Device Lot NumberHCG9030087
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/12/2019
Initial Date FDA Received11/27/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age49 YR
-
-