Model Number NOT APPLICABLE
Device Problems Disconnection (1171); Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/29/2019
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction tubing leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot h333 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot h333 for the reported issue shows no trends.Trends were reviewed for complaint category, tubing leak.No trends were detected for this complaint category.Photographs of the complaint kit were returned for evaluation.A review of the customer provided photographs verify the recirculation pump loop tubing (black stripe tubing) disconnected from the t-connector.The photographs show a blood leak on the instrument pump deck as a result of the tubing becoming disconnected.A material trace of the components used to manufacture kit lot h333 showed no non-conformances.A device history record review did not identify any related nonconformances and this kit lot had passed all lot release testing.The root cause of the tubing becoming disconnected was most likely a weak solvent bond joint between the black stripe tubing and t-connector.The root cause of the leak was most likely a manufacturing operator error during the tube bonding process.Retraining has been completed for all bonding operators.No further action is required at this time.This investigation is now complete.(b)(4).
 
Event Description
The customer contacted mallinckrodt to report they experienced a tubing leak with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported the recirculation pump tubing segment disconnected during the buffy coat collection phase of the procedure.The customer reported approximately 1500 ml of whole blood was processed.The ecp treatment was aborted and no residual blood within the kit was returned to the patient.The customer stated the patient was stable.The customer has returned photographs for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS, INC.
bedminster NJ 07921
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin, D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
1425 us route 206
bedminster, NJ 07921
MDR Report Key9390350
MDR Text Key209846187
Report Number2523595-2019-00141
Device Sequence Number1
Product Code LNR
UDI-Device Identifier20705030200003
UDI-Public(01)20705030200003(10)H333(17)210501
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2021
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXUSA
Device Lot NumberH333
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/29/2019
Initial Date FDA Received11/28/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/28/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age32 YR
-
-