Model Number 10-80-00
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/01/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).Livanova (b)(4) initiated an investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 roller pump was getting a motor controller error during procedure.There was no report of patient injury.
 
Manufacturer Narrative
H.10: a livanova field service representative was dispatched to the customer's facility to investigate the device and could not reproduce the issue.However, he replaced the motor control board as precaution.The pump worked as expected and was put back in service.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich
MDR Report Key9390954
MDR Text Key179414167
Report Number9611109-2019-00926
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/01/2019
Initial Date FDA Received11/29/2019
Supplement Dates Manufacturer Received01/02/2020
Supplement Dates FDA Received01/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-