• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL Back to Search Results
Model Number M00510860
Device Problems Use of Device Problem (1670); Separation Failure (2547)
Patient Problem No Code Available (3191)
Event Date 11/06/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).According to the complainant, the suspect device is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a trapezoid rx basket was used in the pancreatic duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, a trapezoid basket was used in conjunction with an alliance handle in an attempt to crush a 9mm size stone.However, the basket failed to crush the stone and the tip did not detach.The handle was cut from the device and an apc circumferential probe was used to cut the basket wire within the patient's duodenum.The basket remained in the pancreatic duct and a pancreatic stent was placed alongside the basket.The patient was discharged with plans to schedule a spyglass procedure to remove the basket a few weeks later.There were no patient complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.The complainant reported that the trapezoid rx basket was used in the pancreatic duct; however, the directions for use state "caution: trapezoid rx wireguided retrieval basket is not intended for use in the pancreas.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key9395238
MDR Text Key168614232
Report Number3005099803-2019-05827
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296379
UDI-Public08714729296379
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/20/2020
Device Model NumberM00510860
Device Catalogue Number1086
Device Lot Number0024469657
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/06/2019
Initial Date FDA Received12/02/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/21/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient Weight91
-
-