Model Number 1050052 |
Device Problems
Fluid/Blood Leak (1250); Mechanical Problem (1384)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Pma/510k: reported device not marketed in the u.S., however, similar devices or devices that share components, raw materials, process methods or other technological characteristics are registered within the u.S.K111959.If additional information becomes available a follow up report will be submitted.
|
|
Event Description
|
It was reported that there was ampoule leakage.The reporter indicated that two pieces of glue leakage occurred in the first batch of use.The hospital spot checked the remaining batch of glue of the same code, and found that there were several leaks.The event occurred prior to use with no patient harm.
|
|
Manufacturer Narrative
|
Analysis and results: there are previous complaints of this code batch of which we manufactured and distributed in the market (b)(4) units.There are no units in stock in b.Braun surgical warehouse.We have received 30 closed pouches (30 closed ampoules).The ampoules received have been optically evaluated and a defect in the sealing bar in 2 of the 30 ampoules received was found.The leakage of the glue occurs at this point.Reviewed the batch manufacturing record, this product had an incidence not related to this issue and was released fulfilling b.Braun surgical specifications.Final conclusion: taking into account that the results of the samples received do not fulfil b.Braun surgical specifications, we conclude that the complaint is confirmed by evidence of failure in the samples received.Actions on distributed product of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, we have opened a capa in the system in order to determine root cause and actions to correct/prevent this defect to happen.Capa number: ak201730130.
|
|
Search Alerts/Recalls
|