• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US APEX HOLE ELIM POSITIVE STOP; PINNACLE HIP SYSTEM : HIP ACETABULAR AUGMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US APEX HOLE ELIM POSITIVE STOP; PINNACLE HIP SYSTEM : HIP ACETABULAR AUGMENT Back to Search Results
Model Number 1246-03-000
Device Problems Difficult to Insert (1316); Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/05/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.If information is obtained that was not available for the initial meadwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
After the implantation of a pinnacle acetabular component and during the process of insertion, the apex hole eliminator surpassed the apex acetabular hole and remained behind the acetabular cup.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APEX HOLE ELIM POSITIVE STOP
Type of Device
PINNACLE HIP SYSTEM : HIP ACETABULAR AUGMENT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JJM (SUZHOU) LTD. 3006356043
no.299 changyang st
suzhou industrial park
suzhou jiangsu 21512 -6
CH   21512-6
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester, PA 19380-0988
6103142063
MDR Report Key9398198
MDR Text Key184373254
Report Number1818910-2019-119145
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295019688
UDI-Public10603295019688
Combination Product (y/n)N
Reporter Country CodeGR
PMA/PMN Number
K963309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1246-03-000
Device Catalogue Number124603000
Device Lot NumberD19051434
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/05/2019
Initial Date FDA Received12/02/2019
Supplement Dates Manufacturer Received01/21/2020
Supplement Dates FDA Received01/24/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/17/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-