MAUDE Adverse Event Report: EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT WITH TRUWAVE DISPOSABLE PRESSURE TRANSDUCER; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT WITH TRUWAVE DISPOSABLE PRESSURE TRANSDUCER; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number PX260
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Device Displays Incorrect Message (2591)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 11/12/2019
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is anticipated.However, the complaint cannot not be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation results when received.Lot number was not provided, therefore review of the manufacturing records could not be completed.
 
Event Description
It was reported that during use, the disposable pressure transducer did not display correct map (mean arterial pressure) values.The map pressure value displayed was less than 10 mmhg.The expected value was the normal map range in a healthy person.There was no error message or alarm displayed.Patient was not treated according to the incorrect values displayed.Patient demographics are unable to be obtained.
 
Manufacturer Narrative
Investigation found that the potential root cause for this nonconformance could be a supplier material defect.As this defect cannot be generated at our manufacturing process, a supplier notification was made to their personnel about the condition.
 
Manufacturer Narrative
One single dpt kit was returned for examination.The reported event of "couldn't show the correct numbers" was not confirmed due to the dpt did not zero or sense pressure on a pressure monitor.Electrical testing showed that the input impedance met specification, but the output impedance was unstable.A crack was observed on the sensor chip inside of the dpt.The crack cut the sensor chip and affected the output circuit.A n investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.Poor dynamic response can be caused by air bubbles, clotting, and excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.The assembly may be tested for dynamic response by observing the pressure waveform on an oscilloscope or monitor.Bedside determination of the dynamic response of the catheter, monitor, kit and transducer system is done after the system is flushed, attached to the patient, zeroed and calibrated.A square-wave test may be performed by pulling the snap tab device and releasing quickly.Pressure readings can change quickly and dramatically because of loss of proper calibration, loose connection, or air in the system.Pressure readings should correlate with the patient¿s clinical manifestations.In this event, there was no patient compromise noted.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE MONITORING KIT WITH TRUWAVE DISPOSABLE PRESSURE TRANSDUCER
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR 
MDR Report Key9405250
MDR Text Key188712544
Report Number2015691-2019-04463
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
PMA/PMN Number
K925638
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 11/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPX260
Device Catalogue NumberPX260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2019
Initial Date Manufacturer Received 11/12/2019
Initial Date FDA Received12/03/2019
Supplement Dates Manufacturer Received12/12/2019
02/05/2020
07/23/2020
Supplement Dates FDA Received01/06/2020
02/06/2020
01/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-