The device evaluation is anticipated.However, the complaint cannot not be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation results when received.Lot number was not provided, therefore review of the manufacturing records could not be completed.
|
One single dpt kit was returned for examination.The reported event of "couldn't show the correct numbers" was not confirmed due to the dpt did not zero or sense pressure on a pressure monitor.Electrical testing showed that the input impedance met specification, but the output impedance was unstable.A crack was observed on the sensor chip inside of the dpt.The crack cut the sensor chip and affected the output circuit.A n investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.Poor dynamic response can be caused by air bubbles, clotting, and excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.The assembly may be tested for dynamic response by observing the pressure waveform on an oscilloscope or monitor.Bedside determination of the dynamic response of the catheter, monitor, kit and transducer system is done after the system is flushed, attached to the patient, zeroed and calibrated.A square-wave test may be performed by pulling the snap tab device and releasing quickly.Pressure readings can change quickly and dramatically because of loss of proper calibration, loose connection, or air in the system.Pressure readings should correlate with the patient¿s clinical manifestations.In this event, there was no patient compromise noted.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|