• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. CONTROL UNIT DYONICS 25; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. CONTROL UNIT DYONICS 25; ARTHROSCOPE Back to Search Results
Catalog Number 7211010
Device Problems Pressure Problem (3012); Insufficient Information (3190)
Patient Problem Inflammation (1932)
Event Date 10/01/2019
Event Type  malfunction  
Event Description
It was reported that during a shoulder surgery, the pump was not working correctly.It was unable to maintain a flow rate or pressure.The device stopped to work throughout the case for prolonged periods, having a difficultly visualization and need for use of flush function several times.An extra 60-90 minutes were added to the case resulting in extra time of the patient under anesthesia.It resulted in greater than normal swelling in the shoulder as well.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3,h6: the reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of fluctuating pressure could not be reproduced.Product passed functional testing per factory test with no faults or errors.Product passed functional testing during 2 hour burn-in on wet station utilizing low and high pressure and flow settings.Raw and zero transducer readings were normal and well within specs during all functional tests.At no time during functional testing did pump over pressurize, fluctuate, or become intermittent.All functions perform as expected.Pressure and flow readings were normal throughout burn-in on wet station.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTROL UNIT DYONICS 25
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key9406048
MDR Text Key169130344
Report Number1643264-2019-00853
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
PMA/PMN Number
K171794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number7211010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2019
Initial Date Manufacturer Received 11/13/2019
Initial Date FDA Received12/03/2019
Supplement Dates Manufacturer Received01/06/2020
Supplement Dates FDA Received01/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age52 YR
-
-