Common device name - equipment, laboratory, general purpose, labeled or promoted for a specific medical use should additional relevant information become available, a supplemental report will be submitted.
|
The actual sample was received for evaluation.A visual inspection was performed, and it was noted that all components were on the tool as required and had been assembled properly.However, damage was observed on the gears, screws, vial adaptor nest, vial holding jaws, and several pieces of the complaint sample that are attached to the slip clutch.The gears, screws, nests, jaws and mechanical function pieces were observed to be rusted and covered in a dried fluid substance.An actual functional test using vials was not performed; however, the docking tool was tested.It was observed that the gears were stripped thus not allowing force to be applied.The reported condition was verified.The cause of the reported condition could not be determined.The potential root cause of the issue is the product was exposed to a solution containing sodium potentially during use.Since the docking tool was not cleaned (as evident) after use, rust formed on the gears from and parts of the tool, thus causing degradation of the metal and stripping of the gears.Should additional relevant information become available, a supplemental report will be submitted.
|