• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. XPS® SYSTEM NAVIGATED BLADE; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. XPS® SYSTEM NAVIGATED BLADE; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number 1884380EM
Device Problem Vibration (1674)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/14/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
A health care provider (hcp) reported that during endoscopic sinus surgery a blade's tip part was oscillating from the beginning.Re connection was performed several times, but the tip was oscillating when rotating.The procedure was completed with backup product(s) with no delay.There was no patient impact.
 
Manufacturer Narrative
The product analysis result indicates that inner assembly would spin, and middle would index freely by hand with no unusual resistance.There was no damage to the tip however the inner shaft was bent just distal to the inner hub which would have resulted in the wobbling.When viewed under magnification, there was damage to the hubs that is consistent with improper or difficulty loading the device into the handpiece: dimples on the front hub prior to the locking area caused by a misalignment of the handpiece locking mechanism; locking area deformation caused by the back side of the front collet of the handpiece; and deformation of the proximal inner hub chevrons caused by the handpiece drive mechanism.There were no loose components.Even with the deformation of the hubs, functionally the device loaded securely into a handpiece ran at 5,000 rpm in oscillate mode/direction but wobbled as reported.A review of the xpi did not indicate any evidence to point to improper manufacturing and there are checks for damage throughout the process.A bur or blade that is not properly loaded would result in improper support from the handpiece, engagement with the suction seal, drive, and locking mechanism.The complaint was confirmed.In the returned condition, there was an out of specification condition that is likely related to the complaint (due to physical damage).The most likely underlying cause is consistent with, misuse / use error.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® SYSTEM NAVIGATED BLADE
Type of Device
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
christy cain
6743 southpoint drive north
jacksonville, FL 32216
9043328353
MDR Report Key9416428
MDR Text Key169356306
Report Number1045254-2019-00651
Device Sequence Number1
Product Code EQJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K130608
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1884380EM
Device Catalogue Number1884380EM
Device Lot Number0218473300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2019
Initial Date FDA Received12/05/2019
Supplement Dates Manufacturer Received01/23/2020
Supplement Dates FDA Received01/27/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-