Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
A customer reported that the device did not discharge during an emergency resuscitation attempt.The device was report to be in use on a patient, causing a delay in therapy/treatment and will be considered a serious injury.However, no direct adverse event to the patient or user was reported.The customer biomedical engineer reported that the device was evaluated after the event, and found the pads and paddles "in good state." the customer biomedical engineer tested the discharge function and found it to be faulty.The biomedical engineer determined that the faulty discharge was caused by residual gel on the paddles and sockets.The device was cleaned, disinfected, and tested again.The device passed all testing.No ecg strips or case events files were provided to philips for evaluation.The customer biomedical engineer analyzed the event files in detail, and summarized the device behavior during the event as follows: ¿the aforementioned demonstrates that the equipment started to be used at 11:27:06 until 11:39, showing a time lapse of 10 minutes when it is observed that the operator(s) make various modifications in the programming of the defibrillation and ecg reading modes (paddles or derivatives).In addition, the history indicates that various automatic disarmings occurred that the equipment generates for safety reasons, as the adjusted power charge lasts more than 30 seconds without the staff carrying out the corresponding discharge.There is no evidence in the history of any fault that the equipment could have shown in the procedure.¿ philips requested but did not receive additional information related to the event.The device remains in service at the customer site.The customer biomedical engineer recommended that the device be tested daily by hospital staff.The information gathered for this report does not provide evidence of a systemic, design, or labeling problem.No further investigation or action is warranted.
|