• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B.BRAUN SURGICAL SA HISTOACRYL BLUE 0.5ML; TISSUE ADHESIVES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B.BRAUN SURGICAL SA HISTOACRYL BLUE 0.5ML; TISSUE ADHESIVES Back to Search Results
Model Number 1050044
Device Problems Fluid/Blood Leak (1250); Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Pma/510k: reported device not marketed in the u.S., however, similar devices or devices that share components, raw materials, process methods or other technological characteristics are registered within the u.S.K111959.When additional information becomes available a follow up report will be submitted.
 
Event Description
It was reported that there was ampoule leakage.The reporter indicated that the glue was sent to the operating room; when it was opened for use, one piece of glue was found leaking.The glue seal is good, but the glue has leaked.The event occurred prior to use.
 
Manufacturer Narrative
Investigation: samples received: 1 open pouch, unopened ampoule.Analysis and results: there are previous complaints of this code batch of which we manufactured and distributed in the market 4,060 units.There are no units in stock in b.Braun surgical warehouse.We have received one open pouch (closed ampoule) showing ampoule leakage.The ampoule received has been optically evaluated and a defect in the sealing bar of the ampoule was found.The leakage of the glue occurs at this point.Reviewed the batch manufacturing record, this product had an incidence not related to this issue and was released fulfilling b.Braun surgical specifications.Final conclusion: taking into account that the results of the sample received do not fulfil b.Braun surgical specifications, and that there are several previous complaints for the same defect, we conclude that the complaint is confirmed by evidence of failure in the sample received.Actions on distributed product of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.A capa has been initiated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HISTOACRYL BLUE 0.5ML
Type of Device
TISSUE ADHESIVES
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, 08191
SP  08191
MDR Report Key9427006
MDR Text Key172755806
Report Number3003639970-2019-00868
Device Sequence Number1
Product Code MPN
Combination Product (y/n)N
PMA/PMN Number
SEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2020
Device Model Number1050044
Device Catalogue Number1050044
Device Lot Number218364N2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2019
Initial Date Manufacturer Received 11/13/2019
Initial Date FDA Received12/06/2019
Supplement Dates Manufacturer Received12/10/2019
Supplement Dates FDA Received12/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-