The referenced high flow insufflation unit was returned to the service center for evaluation of the reported "pressure settings kept jumping up randomly".A visual inspection was performed and found rust on the bottom chassis and multiple deep scratch marks on top cover.However, the front panel switches and selected settings were operational.The pressure, flow rate, and volume adjusted accordingly.The relief mode, warning tones of excessive pressure, and tube obstruction functioned properly.The warning against excessive peritoneal cavity pressure was working properly as the pinch valve worked appropriately and restored pressure to preset within 10 seconds once over-pressuring was detected.Additionally, the device was left on for an hour, noted the gas supply remained stable, and no air leakage.The pressure settings was also stable, and did not keep jumping up 40 or change the values randomly.However, during a second inspection, the service group found the pressure indicator's two green leds light up at the same time.This condition occurred intermittently due to a faulty main board.However, it does would not contribute to the reported event.A review of the service history revealed no service records for the high flow insufflation unit.The cause of the reported "pressure settings kept jumping up randomly" could not be confirmed as the pressure settings worked appropriately.As a preventive measure, the instruction manual notes, if excessive pressure warning continually sounds, it could be excessive gas volume released from other equipment.Reduce gas volume released from other equipment.Additionally, the instruction manual indicates to, never connect the suction tube and the insufflation tube to the same veress needle or trocar.Doing so will impede measurement of the abdominal pressure which may result in excessive abdominal pressure, or make it impossible to maintain the selected abdominal pressure.If the insufflator emits a warning (warning light or alarm)for intra-abdominal over-pressurization, quickly open the stopcock or valve of the trocar.
|