• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number T-JAPAN
Device Problem Mechanics Altered (2984)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/07/2019
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported a vitrectomy probe did not actuate during a procedure.The condition of aspirating was unknown.The procedure was completed after replacing the product with another one.There was no harm to the patient.
 
Manufacturer Narrative
One opened probe was received with a tip protector for the report of probe did not actuate.The returned sample was visually inspected and found to be non-conforming with the inner cutter in the port.The sample was then functionally tested for actuation, aspiration, and cut and was found to be conforming for all three functional tests.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The returned sample was found to be functionally conforming, therefore probe did not actuate as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No action was taken as the returned probe was functionally conforming.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Any non-conformances found are removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key9433894
MDR Text Key176116195
Report Number2028159-2019-02183
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 03/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberT-JAPAN
Device Catalogue Number8065752044
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2020
Initial Date Manufacturer Received 11/20/2019
Initial Date FDA Received12/06/2019
Supplement Dates Manufacturer Received02/14/2020
Supplement Dates FDA Received03/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CONSTELLATION ULTRAVIT PROBE
-
-