ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problem Insufficient Information (3190)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/14/2019
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of monitor sn (b)(4) has been completed.  upon investigation the monitor failed to power on.  the cause for the failure was isolated to a shorted c1 capacitor on the computer/analog pca. the root cause for the shorted c1 capacitor could not be positively identified.No adverse event resulted from the defective monitor.
 
Event Description
A patient's monitor was returned for an unrelated problem.Upon investigation a reportable malfunction was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
rachel mahoney
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key9438742
MDR Text Key176435945
Report Number3008642652-2019-09657
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005005
UDI-Public00855778005005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 12/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2019
Initial Date FDA Received12/09/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-