• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND, INC TRANSDUCER X8-2T

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND, INC TRANSDUCER X8-2T Back to Search Results
Model Number 989605455171
Device Problem Failure to Sense (1559)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The service engineer tested the system and was able to confirm the reported problem.A visual inspection revealed deposits inside the transducer connector which caused a poor connection between the system and the transducer.After carefully removing the deposits and cleaning the contacts of the connector, the system and transducer passed all tests and resumed normal operation.Both the system and transducer remain in use at the customer site with no return anticipated; therefore, no further failure analysis can be performed.
 
Event Description
A customer reported encountering an incident where their epiq ultrasound system would not recognize the x8-2t tee model transducer during an elective aortic valve replacement.The failure occurred while the transducer was inserted in the patient, but prior to the bypass surgery.After rebooting the system twice, the transducer became fully functional and the procedure was successfully completed with no adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSDUCER X8-2T
Type of Device
TRANSDUCER
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
paul corrigan
22100 bothell everett highway
bothell, WA 98021
4254877000
MDR Report Key9440524
MDR Text Key176315336
Report Number3019216-2019-00087
Device Sequence Number1
Product Code ITX
UDI-Device Identifier00884838069718
UDI-Public(01)00884838069718
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K163120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 11/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number989605455171
Device Lot NumberB2XDCJ
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/12/2019
Initial Date FDA Received12/09/2019
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-